The modular taper junction in total hip replacements has been implicated as a possible source of wear. The finite-element (FE) method can be used to study the wear potential at the taper junction. For such simulations it is important to implement representative contact parameters, in order to achieve accurate results. One of the main parameters in FE simulations is the coefficient of friction. However, in current literature, there is quite a wide spread in coefficient of friction values (0.15 − 0.8), which has a significant effect on the outcome of the FE simulations. Therefore, to obtain more accurate results, one should use a coefficient of friction that is determined for the specific material couple being analyzed. In this study, the static coefficient of friction was determined for two types of titanium-on-titanium stem-adaptor couples, using actual cut-outs of the final implants, to ensure that the coefficient of friction was determined consistently for the actual implant material and surface finish characteristics. Two types of tapers were examined, Biomet type-1 and 12/14, where type-1 has a polished surface finish and the 12/14 is a microgrooved system. We found static coefficients of friction of 0.19 and 0.29 for the 12/14 and type-1 stem-adaptor couples, respectively.
Skip Nav Destination
Article navigation
March 2016
Technical Briefs
Experimental Measurement of the Static Coefficient of Friction at the Ti–Ti Taper Connection in Total Hip Arthroplasty
T. Bitter,
T. Bitter
Orthopaedic Research Laboratory,
Radboud University Medical Centre,
Geert grooteplein-zuid 10 (547),
Nijmegen 6500 HB,
The Netherlands
e-mail: Thom.Bitter@radboudumc.nl
Radboud University Medical Centre,
Geert grooteplein-zuid 10 (547),
Nijmegen 6500 HB,
The Netherlands
e-mail: Thom.Bitter@radboudumc.nl
Search for other works by this author on:
I. Khan,
I. Khan
Biomet UK Healthcare Ltd,
Murdock Road,
Dorcan Industrial Estate,
Swindon SN3 5HY, UK
e-mail: Imran.khan@biomet.com
Murdock Road,
Dorcan Industrial Estate,
Swindon SN3 5HY, UK
e-mail: Imran.khan@biomet.com
Search for other works by this author on:
T. Marriott,
T. Marriott
Biomet UK Healthcare Ltd,
Murdock Road,
Dorcan Industrial Estate,
Swindon SN3 5HY, UK
e-mail: Tim.marriott@biomet.com
Murdock Road,
Dorcan Industrial Estate,
Swindon SN3 5HY, UK
e-mail: Tim.marriott@biomet.com
Search for other works by this author on:
B. W. Schreurs,
B. W. Schreurs
Department of Orthopaedics,
Radboud University Medical Centre,
Nijmegen 6500 HB,
The Netherlands
e-mail: Wim.schreurs@radboudumc.nl
Radboud University Medical Centre,
Nijmegen 6500 HB,
The Netherlands
e-mail: Wim.schreurs@radboudumc.nl
Search for other works by this author on:
N. Verdonschot,
N. Verdonschot
Orthopaedic Research Laboratory,
Radboud University Medical Centre,
Geert grooteplein-zuid 10 (547),
Nijmegen 6500 HB,
The Netherlands
e-mail: Nico.Verdonschot@radboudumc.nl
Radboud University Medical Centre,
Geert grooteplein-zuid 10 (547),
Nijmegen 6500 HB,
The Netherlands
e-mail: Nico.Verdonschot@radboudumc.nl
Search for other works by this author on:
D. Janssen
D. Janssen
Orthopaedic Research Laboratory,
Radboud University Medical Centre,
Geert grooteplein-zuid 10 (547),
Nijmegen 6500 HB,
The Netherlands
e-mail: Dennis.Janssen@radboudumc.nl
Radboud University Medical Centre,
Geert grooteplein-zuid 10 (547),
Nijmegen 6500 HB,
The Netherlands
e-mail: Dennis.Janssen@radboudumc.nl
Search for other works by this author on:
T. Bitter
Orthopaedic Research Laboratory,
Radboud University Medical Centre,
Geert grooteplein-zuid 10 (547),
Nijmegen 6500 HB,
The Netherlands
e-mail: Thom.Bitter@radboudumc.nl
Radboud University Medical Centre,
Geert grooteplein-zuid 10 (547),
Nijmegen 6500 HB,
The Netherlands
e-mail: Thom.Bitter@radboudumc.nl
I. Khan
Biomet UK Healthcare Ltd,
Murdock Road,
Dorcan Industrial Estate,
Swindon SN3 5HY, UK
e-mail: Imran.khan@biomet.com
Murdock Road,
Dorcan Industrial Estate,
Swindon SN3 5HY, UK
e-mail: Imran.khan@biomet.com
T. Marriott
Biomet UK Healthcare Ltd,
Murdock Road,
Dorcan Industrial Estate,
Swindon SN3 5HY, UK
e-mail: Tim.marriott@biomet.com
Murdock Road,
Dorcan Industrial Estate,
Swindon SN3 5HY, UK
e-mail: Tim.marriott@biomet.com
B. W. Schreurs
Department of Orthopaedics,
Radboud University Medical Centre,
Nijmegen 6500 HB,
The Netherlands
e-mail: Wim.schreurs@radboudumc.nl
Radboud University Medical Centre,
Nijmegen 6500 HB,
The Netherlands
e-mail: Wim.schreurs@radboudumc.nl
N. Verdonschot
Orthopaedic Research Laboratory,
Radboud University Medical Centre,
Geert grooteplein-zuid 10 (547),
Nijmegen 6500 HB,
The Netherlands
e-mail: Nico.Verdonschot@radboudumc.nl
Radboud University Medical Centre,
Geert grooteplein-zuid 10 (547),
Nijmegen 6500 HB,
The Netherlands
e-mail: Nico.Verdonschot@radboudumc.nl
D. Janssen
Orthopaedic Research Laboratory,
Radboud University Medical Centre,
Geert grooteplein-zuid 10 (547),
Nijmegen 6500 HB,
The Netherlands
e-mail: Dennis.Janssen@radboudumc.nl
Radboud University Medical Centre,
Geert grooteplein-zuid 10 (547),
Nijmegen 6500 HB,
The Netherlands
e-mail: Dennis.Janssen@radboudumc.nl
1Corresponding author.
Manuscript received July 9, 2015; final manuscript received December 23, 2015; published online January 29, 2016. Assoc. Editor: Michael Detamore.
J Biomech Eng. Mar 2016, 138(3): 034505 (5 pages)
Published Online: January 29, 2016
Article history
Received:
July 9, 2015
Revised:
December 23, 2015
Citation
Bitter, T., Khan, I., Marriott, T., Schreurs, B. W., Verdonschot, N., and Janssen, D. (January 29, 2016). "Experimental Measurement of the Static Coefficient of Friction at the Ti–Ti Taper Connection in Total Hip Arthroplasty." ASME. J Biomech Eng. March 2016; 138(3): 034505. https://doi.org/10.1115/1.4032446
Download citation file:
Get Email Alerts
Cited By
Evaluation of an Unloading Concept for Knee Osteoarthritis: A Pilot Study in a Small Patient Group
J Biomech Eng (January 2024)
Related Articles
Study of Micromotion in Modular Acetabular Components During Gait and Subluxation: A Finite Element Investigation
J Biomech Eng (April,2008)
Specimen-Specific Finite Element Models for Predicting Fretting Wear in Total Hip Arthroplasty Tapers
J Biomech Eng (July,2020)
Tribological and Nanomechanical Properties of Unmodified and Crosslinked Ultra-High Molecular Weight Polyethylene for Total Joint Replacements
J. Tribol (April,2004)
Adaptation of the AnyBody™ Musculoskeletal Shoulder Model to the Nonconforming Total Shoulder Arthroplasty Context
J Biomech Eng (October,2015)
Related Proceedings Papers
Related Chapters
The Tribological Character of LB Films of Dipalmitoylphosphatidylcholine (DPPC)
International Conference on Mechanical and Electrical Technology, 3rd, (ICMET-China 2011), Volumes 1–3
Materials
Design and Application of the Worm Gear
Friction and Wear of Polymers and Composites
Tribology of Mechanical Systems: A Guide to Present and Future Technologies