A new pressure, volume, temperature, and time (PVTt) primary gas flow standard for calibrating flowmeters has an expanded uncertainty of between 0.02% and 0.05%. The standard diverts a steady flow into a collection tank of known volume during a measured time interval. The standard spans the flow range of 1 slm1 to 2000 slm using two collection tanks (34 L and 677 L) and two flow diversion systems. We describe the novel features of the standard and analyze its uncertainty. The thermostatted collection tank allows determination of the average gas temperature to 7 mK (0.0023%) within an equilibration time of 20 min. We developed a mass cancellation procedure that reduced the uncertainty contributions from the inventory volume to 0.017% at the highest flow rate. Flows were independently measured throughout the overlapping flow range of the two systems and they agreed within 0.015 %. The larger collection system was evaluated at high flows by comparing single and double diversions; the maximum difference was 0.0075%.
Skip Nav Destination
Article navigation
November 2003
Technical Papers
Volumetric Gas Flow Standard With Uncertainty of 0.02% to 0.05%
John D. Wright, Project Leader, Fluid Flow Group, Mem. ASME,,
John D. Wright, Project Leader, Fluid Flow Group, Mem. ASME,
Search for other works by this author on:
Michael R. Moldover, Group Leader, Fluid Science Group,,
Michael R. Moldover, Group Leader, Fluid Science Group,
Search for other works by this author on:
Aaron N. Johnson, Fluid Flow Group,
Aaron N. Johnson, Fluid Flow Group
Process Measurements Division, Chemical Science and Technology Laboratory, National Institute of Standards and Technology, 100 Bureau Drive, Gaithersburg, MD 20877
Search for other works by this author on:
Akisato Mizuno, Professor,
Akisato Mizuno, Professor,
Department of Mechanical Engineering, Kogakuin University, Nakano-Machi 2665, Hachioji-shi, Tokyo 192-0015, Japan
Search for other works by this author on:
Aaron N. Johnson, Fluid Flow Group
Process Measurements Division, Chemical Science and Technology Laboratory, National Institute of Standards and Technology, 100 Bureau Drive, Gaithersburg, MD 20877
Akisato Mizuno, Professor,
Department of Mechanical Engineering, Kogakuin University, Nakano-Machi 2665, Hachioji-shi, Tokyo 192-0015, Japan
Contributed by the Fluids Engineering Division for publication in the JOURNAL OF FLUIDS ENGINEERING. Manuscript received by the Fluids Engineering Division Nov. 15, 2002; revised manuscript received June 4, 2003. Associate Editor: S. Ceccio.
J. Fluids Eng. Nov 2003, 125(6): 1058-1066 (9 pages)
Published Online: January 12, 2004
Article history
Received:
November 15, 2002
Revised:
June 4, 2003
Online:
January 12, 2004
Citation
Wright, J. D., Moldover, M. R., Johnson, A. N., and Mizuno, A. (January 12, 2004). "Volumetric Gas Flow Standard With Uncertainty of 0.02% to 0.05% ." ASME. J. Fluids Eng. November 2003; 125(6): 1058–1066. https://doi.org/10.1115/1.1624428
Download citation file:
Get Email Alerts
Cited By
Related Articles
Extrapolation and Curve-Fitting of Calibration Data for Differential Pressure Flow Meters
J. Eng. Gas Turbines Power (February,2010)
Theoretical Analysis of Experimentally Observed Perplexing Calibration Characteristics of Ball-in-Vortex Flow-Meter
J. Fluids Eng (September,2005)
The Systematic Uncertainty of Laboratory Flow Calibrations
J. Fluids Eng (March,1998)
Discharge Coefficient Correlations for Circular-Arc Venturi Flowmeters at Critical (Sonic) Flow
J. Fluids Eng (June,1974)
Related Proceedings Papers
Related Chapters
Pulsation and Vibration Analysis of Compression and Pumping Systems
Pipeline Pumping and Compression System: A Practical Approach, Third Edition
Pulsation and Vibration Analysis of Compression and Pumping Systems
Pipeline Pumping and Compression Systems: A Practical Approach, Second Edition
Compressive Deformation of Hot-Applied Rubberized Asphalt Waterproofing
Roofing Research and Standards Development: 10th Volume